生物安全柜 安全柜
型号:HG06-BHC-1000 ⅡA
洁净等级Cleanliness ISO 5 (10级Class 100)
过滤器级别/过滤效率Filtration Efficiency HEPA/ULPA @0.3-0.12 um 99.995%-99.999%
Aver Velocity平均风速 门内侧Inside the door(m/s) 0.38±0.025m/s
中间 in the middle 0.26±0.025m/s
里侧Inner(m/s) 0.27±0.025m/s
工作窗口吸入风速Velocity inhaled in front ≥0.50±0.025m/s
噪声Noise ≤62dB(A)
振动半峰值Vibration ≤5μm
电源Supply AC,单相220V/50Hz (AC, 220V,1¢,50Hz)
Protectiontesting气流平衡生物防护 人员防护Operator protection 5-8×10 CFU/ml(连续3次3 consecutive times) a.撞击采样器的菌落总数≤10CFU/每次Total
b.狭缝式采样器菌落总数≤5CFU/每次Total
受试产品防护Tested product protection 5-8×10 CFU/ml(连续3次3 consecutive times) 菌落总数≤5CFU/每次Total
交叉感染防护Cross contamination 5-8×10 CFU/ml(连续3次3 consecutive times) 菌落总数≤2CFU/每次Total
zui大功率Max. Consumption (KVA) ≤1.5kw (含备用插座 with spare socket)
重量Mass (kg) 380
工作区尺寸(宽×深×高)Work Dimension (W*D*N) 1000×570×635mm
装置外形尺寸Overal Dimension (mm) 1360×940×2280
荧光灯规格及数量Size and Qty of fluorescent lamp 30W×①
紫外灯规格及数量Size and Qty of UV lamp 30W×①
照度Lumin. >=750LX
排风管口径Size of air exhausted port DN160
出风方向Air Direction 左出、右出、顶出/Left out, Right out, Top out